TY - JOUR TI - Does Electron Spectrum Affect TLD-100 Dose Response in 6 MV Photon Beam Irradiation? AB - In this study, the electron spectrum effect on the TLD-100 dosimeter response to a 6 MV photon beam in different media like water, aluminum, polystyrene, iron, copper, and lead using Monte Carlo and Burlin cavity theory was evaluated. To calculate and compare the dose to medium to dose to cavity correction factors (f), the electronic equilibrium spectrum produced by the 6 MV photon beam and its maximum electron energy in different media were used. The electronic equilibrium spectra were obtained using Beamdp Monte Carlo Simulation. Using two different methods, the cavity theory was applied to obtain the response of the TLD-100 to 6 MV photon beam in the media considered. In the first method, the average mass collision stopping power ratios and the average mass effective attenuation coefficients were calculated using the electron spectrum of 6 MV. In the second method, these parameters were calculated based on the maximum energy value of 6 MV. The maximum difference between the f values obtained using the two methods was about 10 % for lead, while it was less than 2.5 % for other media. Consequently, the differences between f factors calculated using these two methods were insignificant except for lead.  AU - sarigul, neslihan DO - 10.17776/csj.1086868 PY - 2022 JO - Cumhuriyet Science Journal VL - 43 IS - 3 SN - 2587-2680 SP - 504 EP - 509 DB - TRDizin UR - http://search/yayin/detay/1131911 ER -