A case of falsely elevated D-dimer result

Öz:
Objectives: Heterophile antibodies can cause interferences in immunometric assays. While many tests are shown to be affected by interference from heterophilic antibodies, the D-dimer test has rarely been reported to be affected. With this case, we report an elevated D-dimer measurement which was not compatible with the clinical presentation. Case presentation: A 41-year-old patient who was admitted to hospital with heart palpitations had a D-dimer elevation irrelevant to his clinical condition. D-dimer measurements were repeated in new samples directly and after being treated in heterophilic blocking tube with two different reagent lots of a latex-based automated immunoturbidimetric assay and an immunoturbidometric assay. D-dimer values were normalized (0–0.5 mg/L) when we used a new lot of reagent on the same instrument or measured by an immunoturbidometric method on the chemistry analyzer. After treatment with HBT, all samples revealed D-dimer results within the reference ranges. Conclusions: The presence of heterophile antibodies in a sample should be considered when an elevated D-dimer value that is not compatible with the clinical presentation is encountered. Apart from the patient’s HA’s causing false results, sporadic susceptibility of the reagents should also be kept in mind as a possibility.
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Olgu Sunumu Erişim Türü: Erişime Açık
  • 1. Weitz JI, Fredenburgh JC, Eikelboom JW. A test in context: D-dimer. J Am Coll Cardiol 2017;70:2411–20.
  • 2. Johnson ED, Schell JC, Rodgers GM. The D-dimer assay. Am J Hematol 2019;94:833–9.
  • 3. Favressea J, Lippib G, Roy PM, Chatelaina B, Jacqmina H, Cate H, et al. Preanalytical, analytical, post-analytical variables, and clinical applications. Crit Rev Clin Lab Sci 2018;55: 548–77.
  • 4. WuY, Xiao YX, Huang TY, Zhang XY, Zhou HB, Zhang XX, et al. What makes D-dimer assays suspicious–heterophilic antibodies? J Clin Lab Anal 2019;33:22687.
  • 5. Ramzi M, Esteve F, Grimaux M. D-dimer measurements and Lipemia Interference of STA-Liatest D-Di on diagnostica stago analyzers. A new approach for interference testing and lipid removal by high-speed centrifugation. Diagnostica Stago; 2018: 7 p. Report No. 301478.
  • 6. Çokluk E, Tuncer FB, Şekeroğlu MR, Gözükara C, Ağaç S, Gözükara Sİ. Effect of heterophyll antibody interference on CA 19-9 measurement: a case report. Sakarya Tip Dergisi 2019;9:558–64.
  • 7. Sun HX, Ge H, Xu ZQ, Sheng HM. Clinical laboratory investigation of a patient with an extremely high D-dimer level: a case report. World J Clin Cases 2020;8:3560–6.
  • 8. Bolstad N, Warren DJ, Nusta K. Heterophilic antibody interference in immunometric assays. Best Pract Res Clin Endocrinol Metab 2013;27:647–61.
  • 9. STA®-Liatest® D-Di Plus, less interferences for even more confidence. Available from: https://www.stago.com/productsservices/ new-products/detail/article.
  • 10. Levinson SS, Miller JJJ. Towards a better understanding of heterophile (and the like) antibody interference with modern immunoassays. Clin Chim Acta 2002;325:1–15.
APA çevlik t, TURKAL R, Haklar G, Sirikci O (2022). A case of falsely elevated D-dimer result. , 686 - 689. 10.1515/tjb-2021-0262
Chicago çevlik tülay,TURKAL RANA,Haklar Goncagul,Sirikci Onder A case of falsely elevated D-dimer result. (2022): 686 - 689. 10.1515/tjb-2021-0262
MLA çevlik tülay,TURKAL RANA,Haklar Goncagul,Sirikci Onder A case of falsely elevated D-dimer result. , 2022, ss.686 - 689. 10.1515/tjb-2021-0262
AMA çevlik t,TURKAL R,Haklar G,Sirikci O A case of falsely elevated D-dimer result. . 2022; 686 - 689. 10.1515/tjb-2021-0262
Vancouver çevlik t,TURKAL R,Haklar G,Sirikci O A case of falsely elevated D-dimer result. . 2022; 686 - 689. 10.1515/tjb-2021-0262
IEEE çevlik t,TURKAL R,Haklar G,Sirikci O "A case of falsely elevated D-dimer result." , ss.686 - 689, 2022. 10.1515/tjb-2021-0262
ISNAD çevlik, tülay vd. "A case of falsely elevated D-dimer result". (2022), 686-689. https://doi.org/10.1515/tjb-2021-0262
APA çevlik t, TURKAL R, Haklar G, Sirikci O (2022). A case of falsely elevated D-dimer result. Türk Biyokimya Dergisi, 47(5), 686 - 689. 10.1515/tjb-2021-0262
Chicago çevlik tülay,TURKAL RANA,Haklar Goncagul,Sirikci Onder A case of falsely elevated D-dimer result. Türk Biyokimya Dergisi 47, no.5 (2022): 686 - 689. 10.1515/tjb-2021-0262
MLA çevlik tülay,TURKAL RANA,Haklar Goncagul,Sirikci Onder A case of falsely elevated D-dimer result. Türk Biyokimya Dergisi, vol.47, no.5, 2022, ss.686 - 689. 10.1515/tjb-2021-0262
AMA çevlik t,TURKAL R,Haklar G,Sirikci O A case of falsely elevated D-dimer result. Türk Biyokimya Dergisi. 2022; 47(5): 686 - 689. 10.1515/tjb-2021-0262
Vancouver çevlik t,TURKAL R,Haklar G,Sirikci O A case of falsely elevated D-dimer result. Türk Biyokimya Dergisi. 2022; 47(5): 686 - 689. 10.1515/tjb-2021-0262
IEEE çevlik t,TURKAL R,Haklar G,Sirikci O "A case of falsely elevated D-dimer result." Türk Biyokimya Dergisi, 47, ss.686 - 689, 2022. 10.1515/tjb-2021-0262
ISNAD çevlik, tülay vd. "A case of falsely elevated D-dimer result". Türk Biyokimya Dergisi 47/5 (2022), 686-689. https://doi.org/10.1515/tjb-2021-0262