TY - JOUR TI - A case of falsely elevated D-dimer result AB - Objectives: Heterophile antibodies can cause interferences in immunometric assays. While many tests are shown to be affected by interference from heterophilic antibodies, the D-dimer test has rarely been reported to be affected. With this case, we report an elevated D-dimer measurement which was not compatible with the clinical presentation. Case presentation: A 41-year-old patient who was admitted to hospital with heart palpitations had a D-dimer elevation irrelevant to his clinical condition. D-dimer measurements were repeated in new samples directly and after being treated in heterophilic blocking tube with two different reagent lots of a latex-based automated immunoturbidimetric assay and an immunoturbidometric assay. D-dimer values were normalized (0–0.5 mg/L) when we used a new lot of reagent on the same instrument or measured by an immunoturbidometric method on the chemistry analyzer. After treatment with HBT, all samples revealed D-dimer results within the reference ranges. Conclusions: The presence of heterophile antibodies in a sample should be considered when an elevated D-dimer value that is not compatible with the clinical presentation is encountered. Apart from the patient’s HA’s causing false results, sporadic susceptibility of the reagents should also be kept in mind as a possibility. AU - Haklar, Goncagul AU - çevlik, tülay AU - TURKAL, RANA AU - Sirikci, Onder DO - 10.1515/tjb-2021-0262 PY - 2022 JO - Türk Biyokimya Dergisi VL - 47 IS - 5 SN - 1303-829X SP - 686 EP - 689 DB - TRDizin UR - http://search/yayin/detay/1167990 ER -