TY - JOUR TI - Dose Calibrator Measurements in the Case of Voltage Fluctuations AB - Radiation dose measurements of radiopharmaceuticals and measurements of dose calibrators at different voltages and fluctuations were evaluated. Low and high doses of technetium, iodine and fluorine pharmaceuticals were measured in different voltages. The results showed that only low doses could make small and insignificant differences. AU - günay, osman AU - Altıntaş, İrem Nur AU - Demir, Mustafa AU - Yeyin, Nami DO - 10.22399/ijcesen.1303582 PY - 2023 JO - International Journal of Computational and Experimental Science and Engineering VL - 9 IS - 2 SN - 2149-9144 SP - 161 EP - 164 DB - TRDizin UR - http://search/yayin/detay/1182934 ER -