Yıl: 2019 Cilt: 30 Sayı: 2 Sayfa Aralığı: 137 - 142 Metin Dili: İngilizce DOI: 10.5606/ehc.2019.64940 İndeks Tarihi: 25-11-2019

Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones

Öz:
Objectives: This study aims to investigate the fitness of two anatomic distal femoral plates with cadaveric femurs and to show whether current plates optimally match each femur. Materials and methods: Two different sets of plates with five, seven and nine shaft holes were applicated on 62 cadaveric femurs. Ball clay was molded onto the entire inner surfaces of the plates and then the plates were fixed to the bones using two self-locking nylon cable zip ties. The volume of ball clay sandwiched in between the plate and bone was calculated and used as a quantitative fit parameter. Data of each plate were analyzed separately. Results: Using Double Medical Technology IncorporatedTM plates, the mean plate to bone volumes were calculated as 8.4 mL (range, 5-14 mL), 10.0 mL (range, 6-17 mL), and 13.1 mL (range, 7-25 mL) in five, seven and nine-hole plates, respectively. Using Zimmer Biomet IncorporatedTM plates, the mean volumes were 10.5 mL (range, 6-21 mL), 12.7 mL (range, 7-22 mL) and 16.3 mL (range, 8-30 mL) in five, seven and nine-hole plates, respectively. Within each group, the measurements were significantly correlated positively with femoral length. Conclusion: Optimal fit may not be achieved in each femur using current distal femoral plate implant sets. Thus additional sizes of plates should be supplied in the implant sets.
Anahtar Kelime:

Konular: Romatoloji Cerrahi Ortopedi

İki farklı anatomik şekillendirilmiş distal femur plak setinin kadavra kemikleriyle uyumunun değerlendirilmesi

Öz:
Amaç: Bu çalışmada iki anatomik distal femur plağının kadavra femurlarıyla uyumluluğu araştırıldı ve mevcut plakların her femura en iyi şekilde uyup uymadığı gösterildi. Gereç ve yöntemler: Gövdesinde beş, yedi ve dokuz deliği olan iki farklı plak seti 62 kadavra femuruna uygulandı. Plakların tüm iç yüzeyleri seramik çamuruyla sıvandı, sonra plaklar iki adet kendinden kilitli naylon kablo bağı kullanılarak kemiklere sabitlendi. Plak ve kemik arasında sıkıştırılan seramik çamurunun hacmi hesaplanarak nicel uyum parametresi olarak kullanıldı. Her plağın verileri ayrı ayrı analiz edildi. Bulgular: Double Medical Technology IncorporatedTM plakları kullanıldığında, plak ile kemik arasındaki hacim ortalamaları beş, yedi ve dokuz delikli plaklarda sırasıyla 8.4 mL (dağılım, 5-14 mL), 10.0 mL (dağılım, 6-17 mL) ve 13.1 mL (dağılım, 7-25 mL) olarak hesaplandı. Zimmer Biomet IncorporatedTM plakları kullanıldığında, hacim ortalamaları beş, yedi ve dokuz delikli plaklarda sırasıyla 10.5 mL (dağılım, 6-21 mL), 12.7 mL (dağılım, 7-22 mL) ve 16.3 mL (dağılım, 8-30 mL) idi. Her grup içinde, ölçümler femur uzunluğu ile pozitif olarak anlamlı şekilde bağıntılı idi. Sonuç: Mevcut distal femur plağı implant setleri kullanılarak her femurda en iyi uyum elde edilemeyebilir. Bu nedenle, implant setlerinde ilave boyutlarda plaklar sağlanmalıdır.
Anahtar Kelime:

Konular: Romatoloji Cerrahi Ortopedi
Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • Gill S, Mittal A, Raj M, Singh P, Singh J, Kumar S. Extra Articular Supracondylar Femur Fractures Managed with Locked Distal Femoral Plate or Supracondylar Nailing: A Comparative Outcome Study. J Clin Diagn Res 2017;11:19-23.
  • Piétu G, Ehlinger M. Minimally invasive internal fixation of distal femur fractures. Orthop Traumatol Surg Res 2017;103:S161-9.
  • Batista BB, Salim R, Paccola CA, Kfuri Junior M. Internal fixators: a safe option for managing distal femur fractures? Acta Ortop Bras 2014;22:159-62.
  • Bel JC. Pitfalls and limits of locking plates. Orthop Traumatol Surg Res 2019;105:103-9.
  • Pesantez R. Surgical reduction. In: Buckley RE, Moran GM, Apivatthakakul T, editors. AO principles of fracture mamagement. 3rd ed. Stuttgart; Thieme Verlag; 2018. p. 117-36.
  • Ahmad M, Nanda R, Bajwa AS, Candal-Couto J, Green S, Hui AC. Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability? Injury 2007;38:358-64.
  • Stoffel K, Dieter U, Stachowiak G, Gächter A, Kuster MS. Biomechanical testing of the LCP--how can stability in locked internal fixators be controlled? Injury 2003;34:11-9.
  • Chen X, He K, Chen Z, Wang L. A parametric approach to construct femur models and their fixation plates. Biotechnol Biotec Eq 2016;30:529-37.
  • Chen X, He K, Chen Z. A novel computer-aided approach for parametric investigation of custom design of fracture fixation plates. Comput Math Methods Med 2017;2017:7372496.
  • Wang, L, He, K, Chen, Z, Yang, Y. A design method for orthopedic plates based on surface features. J Mech Design 2017;139:024502.
  • Hwang JH, Oh JK, Oh CW, Yoon YC, Choi HW. Mismatch of anatomically pre-shaped locking plate on Asian femurs could lead to malalignment in the minimally invasive plating of distal femoral fractures: a cadaveric study. Arch Orthop Trauma Surg 2012;132:51-6.
  • Pramod G, Vijayakumar AV. Study on the complications of locking compression plate used for fracture lower end of femur. Int J Orthop Sci 2016;2:146-9.
  • Goyal KS, Skalak AS, Marcus RE, Vallier HA, Cooperman DR. Analysis of anatomic periarticular tibial plate fit on normal adults. Clin Orthop Relat Res 2007;461:245-57.
  • Hazreen H, Beat S, Michael S, Yuan Tong G, Prasad KY. Quantitative fit assessment of a precontoured fracture fixation plate: Its automation and an investigation on the borderline cases. Adv Mater Res 2011;339:685-9.
  • Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg [Am] 2007;89:2260-5.
  • Kozic N, Weber S, Büchler P, Lutz C, Reimers N, González Ballester MA, et al. Optimisation of orthopaedic implant design using statistical shape space analysis based on level sets. Med Image Anal 2010;14:265-75.
  • Park AY, DiStefano JG, Nguyen TQ, Buckley JM, Montgomery WH, et al. Congruency of scapula locking plates: implications for implant design. Am J Orthop (Belle Mead NJ) 2012;41:E53-6.
  • Schulz AP, Reimers N, Wipf F, Vallotton M, Bonaretti S, Kozic N, et al. Evidence Based Development of a Novel Lateral Fibula Plate (VariAx Fibula) Using a Real CT Bone Data Based Optimization Process During Device Development. Open Orthop J 2012;6:1-7.
  • Schmutz B, Wullschleger ME, Kim H, Noser H, Schütz MA. Fit assessment of anatomic plates for the distal medial tibia. J Orthop Trauma 2008;22:258-63.
  • Atik OŞ. Which articles do we prefer to publish? Eklem Hastalik Cerrahisi 2018;29:1.
  • Wang L, He K, Chen Z. Statistical analyses of femur parameters for designing anatomical plates. Comput Math Methods Med 2016;2016:1247560.
  • Giannicola G, Manauzzi E, Sacchetti FM, Greco A, Bullitta G, Vestri A, et al. The applicability of the Mayo Clinic congruent radial head plate: cadaveric study. Musculoskelet Surg 2011;95:1-5.
  • Giannicola G, Manauzzi E, Sacchetti FM, Greco A, Bullitta G, Vestri A, et al. Anatomical variations of the proximal radius and their effects on osteosynthesis. J Hand Surg [Am] 2012;37:1015-23.
APA Utkan A, İğdir V, ARSLAN A, ŞEN ESMER T, ÖZKURT B (2019). Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. , 137 - 142. 10.5606/ehc.2019.64940
Chicago Utkan Ali,İğdir Volkan,ARSLAN Aydın,ŞEN ESMER TÜLİN,ÖZKURT Bülent Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. (2019): 137 - 142. 10.5606/ehc.2019.64940
MLA Utkan Ali,İğdir Volkan,ARSLAN Aydın,ŞEN ESMER TÜLİN,ÖZKURT Bülent Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. , 2019, ss.137 - 142. 10.5606/ehc.2019.64940
AMA Utkan A,İğdir V,ARSLAN A,ŞEN ESMER T,ÖZKURT B Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. . 2019; 137 - 142. 10.5606/ehc.2019.64940
Vancouver Utkan A,İğdir V,ARSLAN A,ŞEN ESMER T,ÖZKURT B Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. . 2019; 137 - 142. 10.5606/ehc.2019.64940
IEEE Utkan A,İğdir V,ARSLAN A,ŞEN ESMER T,ÖZKURT B "Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones." , ss.137 - 142, 2019. 10.5606/ehc.2019.64940
ISNAD Utkan, Ali vd. "Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones". (2019), 137-142. https://doi.org/10.5606/ehc.2019.64940
APA Utkan A, İğdir V, ARSLAN A, ŞEN ESMER T, ÖZKURT B (2019). Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.), 30(2), 137 - 142. 10.5606/ehc.2019.64940
Chicago Utkan Ali,İğdir Volkan,ARSLAN Aydın,ŞEN ESMER TÜLİN,ÖZKURT Bülent Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.) 30, no.2 (2019): 137 - 142. 10.5606/ehc.2019.64940
MLA Utkan Ali,İğdir Volkan,ARSLAN Aydın,ŞEN ESMER TÜLİN,ÖZKURT Bülent Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.), vol.30, no.2, 2019, ss.137 - 142. 10.5606/ehc.2019.64940
AMA Utkan A,İğdir V,ARSLAN A,ŞEN ESMER T,ÖZKURT B Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.). 2019; 30(2): 137 - 142. 10.5606/ehc.2019.64940
Vancouver Utkan A,İğdir V,ARSLAN A,ŞEN ESMER T,ÖZKURT B Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones. Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.). 2019; 30(2): 137 - 142. 10.5606/ehc.2019.64940
IEEE Utkan A,İğdir V,ARSLAN A,ŞEN ESMER T,ÖZKURT B "Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones." Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.), 30, ss.137 - 142, 2019. 10.5606/ehc.2019.64940
ISNAD Utkan, Ali vd. "Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones". Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.) 30/2 (2019), 137-142. https://doi.org/10.5606/ehc.2019.64940