Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report

Öz:
The aim of this case report is to present an unusual mechanism of failure of a dynamic locking plate (DLP), previously reported in dual lag screw proximal femoral nails. A 78-year-old female patient experienced a displaced right femoral neck fracture (FNF) which was managed with DLP. At the postoperative sixth-week appointment, her hip radiographs showed that two of three telescoping screws had slid within the telescrew barrel, and one screw was fully extended and cutting through the femoral head. At the third-month checkup, radiographs revealed an unhealed FNF and 12 mm sliding on all three telescoping screws. At the f i na l 11th-month follow-up, two screws were completely collapsed and the previously nonpenetrating screw was partially extended and cutting through the femoral head. Femoral neck shortening and femoral neck nonunion were noticed as well. In conclusion, the postoperative course of the case supports the hypothesis of a specific pattern of DLP failure resembling reversed Z- and Z-effects found in dual lag screw proximal femoral nails.
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Olgu Sunumu Erişim Türü: Erişime Açık
  • 1. Fixation using Alternative Implants for the Treatment of Hip fractures (FAITH) Investigators. Fracture fixation in the operative management of hip fractures (FAITH): an international, multicentre, randomised controlled trial. Lancet 2017;389:1519-27.
  • 2. Şahin A, Agar A, Gülabi D, Ertürk C. Comparison of dynamic hip screw and antirotation screw with cannulated screw in the treatment of transcervical collum femoris fractures. Jt Dis Relat Surg 2020;31:320-7.
  • 3. Weil YA, Qawasmi F, Liebergall M, Mosheiff R, Khoury A. Use of fully threaded cannulated screws decreases femoral neck shortening after fixation of femoral neck fractures. Arch Orthop Trauma Surg 2018;138:661-7.
  • 4. Parker MJ, Stedtfeld HW. Internal fixation of intracapsular hip fractures with a dynamic locking plate: initial experience and results for 83 patients treated with a new implant. Injury 2010;41:348-51.
  • 5. Parker MJ, Raghavan R, Gurusamy K. Incidence of fracturehealing complications after femoral neck fractures. Clin Orthop Relat Res 2007;458:175-9.
  • 6. Yin H, Pan Z, Jiang H. Is dynamic locking plate(Targon FN) a better choice for treating of intracapsular hip fracture? A meta-analysis. Int J Surg 2018;52:30-34.
  • 7. Biber R, Brem M, Bail HJ. Targon Femoral Neck for femoralneck fracture fixation: lessons learnt from a series of one hundred and thirty five consecutive cases. Int Orthop 2014;38:595-9.
  • 8. Xiong WF, Chang SM, Zhang YQ, Hu SJ, Du SC. Inferior calcar buttress reduction pattern for displaced femoral neck fractures in young adults: a preliminary report and an effective alternative. J Orthop Surg Res 2019;14:70.
  • 9. Parker M, Cawley S, Palial V. Internal fixation of intracapsular fractures of the hip using a dynamic locking plate: Two-year follow-up of 320 patients. Bone Joint J 2013;95-B:1402-5.
  • 10. Eschler A, Brandt S, Gierer P, Mittlmeier T, Gradl G. Angular stable multiple screw fixation (Targon FN) versus standard SHS for the fixation of femoral neck fractures. Injury 2014;45 Suppl 1:S76-80.
  • 11. Körver RJ, Wieland AW, Kaarsemaker S, Nieuwenhuis JJ, Janzing HM. Clinical experience, primary results and pitfalls in the treatment of intracapsular hip fractures with the Targon® FN locking plate. Injury 2013;44:1926-9.
  • 12. Werner-Tutschku W, Lajtai G, Schmiedhuber G, Lang T, Pirkl C, Orthner E. Intra- and perioperative complications in the stabilization of per- and subtrochanteric femoral fractures by means of PFN. Unfallchirurg 2002;105:881-5.
  • 13. Singh M, Nagrath AR, Maini PS. Changes in trabecular pattern of the upper end of the femur as an index of osteoporosis. J Bone Joint Surg Am 1970;52:457-67.
  • 14. Weil YA, Gardner MJ, Mikhail G, Pierson G, Helfet DL, Lorich DG. Medial migration of intramedullary hip fixation devices: a biomechanical analysis. Arch Orthop Trauma Surg 2008;128:227-34.
  • 15. Boldin C, Seibert FJ, Fankhauser F, Peicha G, Grechenig W, Szyszkowitz R. The proximal femoral nail (PFN)--a minimal invasive treatment of unstable proximal femoral fractures: a prospective study of 55 patients with a followup of 15 months. Acta Orthop Scand 2003;74:53-8.
  • 16. Strauss EJ, Kummer FJ, Koval KJ, Egol KA. The “Z-effect“ phenomenon defined: a laboratory study. J Orthop Res 2007;25:1568-73.
  • 17. Atik OŞ. Is there something new and interesting in my article? Eklem Hastalik Cerrahisi 2019;30:69.
APA Morochovič R, Kňazovický M, Paulo M, Takacova K, Burda R (2021). Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. , 249 - 252. 10.5606/ehc.2021.77186
Chicago Morochovič Radoslav,Kňazovický Miloš,Paulo Martin,Takacova Katarina,Burda Rastislav Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. (2021): 249 - 252. 10.5606/ehc.2021.77186
MLA Morochovič Radoslav,Kňazovický Miloš,Paulo Martin,Takacova Katarina,Burda Rastislav Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. , 2021, ss.249 - 252. 10.5606/ehc.2021.77186
AMA Morochovič R,Kňazovický M,Paulo M,Takacova K,Burda R Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. . 2021; 249 - 252. 10.5606/ehc.2021.77186
Vancouver Morochovič R,Kňazovický M,Paulo M,Takacova K,Burda R Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. . 2021; 249 - 252. 10.5606/ehc.2021.77186
IEEE Morochovič R,Kňazovický M,Paulo M,Takacova K,Burda R "Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report." , ss.249 - 252, 2021. 10.5606/ehc.2021.77186
ISNAD Morochovič, Radoslav vd. "Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report". (2021), 249-252. https://doi.org/10.5606/ehc.2021.77186
APA Morochovič R, Kňazovický M, Paulo M, Takacova K, Burda R (2021). Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. Joint diseases and related surgery, 32(1), 249 - 252. 10.5606/ehc.2021.77186
Chicago Morochovič Radoslav,Kňazovický Miloš,Paulo Martin,Takacova Katarina,Burda Rastislav Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. Joint diseases and related surgery 32, no.1 (2021): 249 - 252. 10.5606/ehc.2021.77186
MLA Morochovič Radoslav,Kňazovický Miloš,Paulo Martin,Takacova Katarina,Burda Rastislav Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. Joint diseases and related surgery, vol.32, no.1, 2021, ss.249 - 252. 10.5606/ehc.2021.77186
AMA Morochovič R,Kňazovický M,Paulo M,Takacova K,Burda R Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. Joint diseases and related surgery. 2021; 32(1): 249 - 252. 10.5606/ehc.2021.77186
Vancouver Morochovič R,Kňazovický M,Paulo M,Takacova K,Burda R Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report. Joint diseases and related surgery. 2021; 32(1): 249 - 252. 10.5606/ehc.2021.77186
IEEE Morochovič R,Kňazovický M,Paulo M,Takacova K,Burda R "Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report." Joint diseases and related surgery, 32, ss.249 - 252, 2021. 10.5606/ehc.2021.77186
ISNAD Morochovič, Radoslav vd. "Reversed Z-effect and Z-effect phenomena in femoral neck fracture treated with a dynamic locking plate fixation: Case report". Joint diseases and related surgery 32/1 (2021), 249-252. https://doi.org/10.5606/ehc.2021.77186