Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants

Öz:
Objective: The study aimed to compare and evaluate the accuracy of iPex, Root ZX mini, and Epex Pro Electronic apex locators (EALs) in diagnosing root perforations in both dry and in different wet conditions: 5% sodium hypochlorite (NaOCl), 2% chlorhexidine (CHX), and 17% Ethylenediaminetetraacetic acid (EDTA). Methods: Thirty extracted, human single rooted mandibular premolars were artificially perforated with a diameter of 1.5 mm in middle third of root. Actual canal lengths (ALs) in millimetre (mm) were evaluated for all teeth up to perforation location, and alginate mould were used to embed the teeth. After this, the electronic measurements were calculated by all EALs up to perforation site using a 20 K-file in both dry and wet canal conditions. Up to the perforation sites, the ALs were subtracted from the electronic length. Statistical analyses were done using One-way ANOVA with post hoc tukey’s test for pairwise comparison and the level of significance was set at 0.05. Results: All three EAL’s detected canal perforations which were clinically acceptable. There was significant difference for dry and wet conditions. Most accurate measurement were seen in dry canals for all three EALs. Root ZX mini in dry condition showed most accurate reading and there was a significant difference when compared with other groups. No significance difference was observed in iPex and Epex Pro Apex locator, and between NaOCl and CHX, CHX and EDTA. Conclusion: Perforations were determined within a clinical acceptable range of 0.03–0.05 mm by all three EALs. Root ZX mini in dry canals gave most accurate measurement. The presence of irrigating solution influenced the accuracy of all the apex locators.
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • 1. Alhadainy HA. Root perforations. A review of literature. Oral Surg Oral Med Oral Pathol 1994; 78(3):368–74.
  • 2. Srivastava V, Jain N, Bagchi S, Negi MPS. Evaluation of the Use of Sixth Generation Apex Locators as a Diagnostic Tool to Detect Root Perforations. Int J Dent Med Spec 2015; 2(4):10 14.
  • 3. Zmener O, Grimberg F, Banegas G, Chiacchio L. Detection and measurement of endodontic root perforations using a newly designed apex-locating handpiece. Endod Dent Traumatol 1999; 15(4):182–5.
  • 4. Fuss Z, Assooline LS, Kaufman AY. Determination of location of root perforations by electronic apex locators. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996; 82(3):324–9.
  • 5. Altunbaş D, Kuştarcı A, Toyoğlu M. The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations. J Endod 2017; 43(3):439–42.
  • 6. Nahmias Y, Aurelio JA, Gerstein H. Expanded use of the electronic canal length measuring devices. J Endod 1983; 9(8):347–9.
  • 7. Mosleh H, Khazaei S, Razavian H, Vali A, Ziaei F. Electronic apex locator: A comprehensive literature review – Part I: Different generations, comparison with other techniques and different usages. Dental Hypotheses 2014; 5(3):84-97.
  • 8. Sunada I. New method for measuring the length of the root canal. J Dent Res 1962; 41(2):375–87.
  • 9. Alothmani OS. The accuracy of Root ZX electronic apex locator. Saudi Endod J 2012; 2(3):115–30.
  • 10. de Vasconcelos BC, do Vale TM, de Menezes AS, Pinheiro-Junior EC, Vivacqua-Gomes N, Bernardes RA, et al. An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010; 110(2):e57–61.
  • 11. D'Assunção FL, Sousa JC, Felinto KC, de Medeiros TC, Leite DT, de Lucena RB, et al. Accuracy and repeatability of 3 apex locators in locating root canal perforations: an ex vivo study. J Endod 2014; 40(8):1241–4.
  • 12. Stöber EK, Duran-Sindreu F, Mercadé M, Vera J, Bueno R, Roig M. An evaluation of root ZX and iPex apex locators: an in vivo study. J Endod 2011; 37(5):608–10.
  • 13. Tsesis I, Rosenberg E, Faivishevsky V, Kfir A, Katz M, Rosen E. Prevalence and associated periodontal status of teeth with root perforation: a retrospective study of 2,002 patients' medical records. J Endod 2010; 36(5):797–800.
  • 14. Saraswathi V, Kedia A, Purayil TP, Ballal V, Saini A. Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study. J Conserv Dent 2016; 19(5):402–5.
  • 15. Ebrahim AK, Wadachi R, Suda H. Accuracy of three different electronic apex locators in detecting simulated horizontal and vertical root fractures. Aust Endod J 2006; 32(2):64–9.
  • 16. Vieyra JP, Acosta J. Comparison of working length determination with radiographs and four electronic apex locators. Int Endod J 2011; 44(6):510–8.
  • 17. Altunbas D, Kustarci A, Arslan D, Er K. In vitro comparison of four different electronic apex locators to determine the major foramen using the clearing technique. Niger J Clin Pract 2014; 17(6):706–10.
  • 18. Baldi JV, Victorino FR, Bernardes RA, de Moraes IG, Bramante CM, Garcia RB, et al. Influence of embedding media on the assessment of electronic apex locators. J Endod 2007; 33(4):476–9.
  • 19. Shin HS, Yang WK, Kim MR, Ko HJ, Cho KM, Park SH, et al. Accuracy of Root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant. Restor Dent Endod 2012; 37(3):149–54.
  • 20. Marroquín BB, Fernández CC, Schmidtmann I, Willershausen B, Goldberg F. Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo study. Head Face Med 2014;10:57.
  • 21. Li YH, Zhou Z, Zheng YQ, Gan N, Tang YY, Li R, et al. Accuracy of three different electronic apex locators in determination of perforation with various conditions in vitro. [Article in Chinese]. Hua Xi Kou Qiang Yi Xue Za Zhi 2011; 29(3):272–5.
  • 22. Duran-Sindreu F, Gomes S, Stöber E, Mercadé M, Jané L, Roig M. In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants. Int Endod J 2013; 46(8):769–74.
  • 23. Venturi M, Breschi L. A comparison between two electronic apex locators: an ex vivo investigation. Int Endod J 2007; 40(5):362–73.
  • 24. Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex locator: an in vitro study. Int Endod J 2002; 35(2):186–92.
  • 25. Çalışkan MK, Kaval ME, Tekin U. Clinical accuracy of two electronic apex locators in teeth with large periapical lesions. Int Endod J 2014; 47(10):920–5.
  • 26. Khandewal D, Ballal NV, Saraswathi MV. Comparative evaluation of accuracy of 2 electronic Apex locators with conventional radiography: an ex vivo study. J Endod 2015; 41(2):201–4.
  • 27. Herrera M, Ábalos C, Lucena C, Jiménez-Planas A, Llamas R. Critical diameter of apical foramen and of file size using the Root ZX apex locator: an in vitro study. J Endod 2011; 37(9):1306–9.
APA bilaiya s, Patni P, Jain P, Pandey S, Raghuwanshi S, bagulkar b (2020). Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. , 6 - 9. 10.14744/eej.2019.07279
Chicago bilaiya sakshi,Patni Pallav,Jain Pradeep,Pandey Sanket Hans,Raghuwanshi Swadhin,bagulkar bhupesh Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. (2020): 6 - 9. 10.14744/eej.2019.07279
MLA bilaiya sakshi,Patni Pallav,Jain Pradeep,Pandey Sanket Hans,Raghuwanshi Swadhin,bagulkar bhupesh Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. , 2020, ss.6 - 9. 10.14744/eej.2019.07279
AMA bilaiya s,Patni P,Jain P,Pandey S,Raghuwanshi S,bagulkar b Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. . 2020; 6 - 9. 10.14744/eej.2019.07279
Vancouver bilaiya s,Patni P,Jain P,Pandey S,Raghuwanshi S,bagulkar b Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. . 2020; 6 - 9. 10.14744/eej.2019.07279
IEEE bilaiya s,Patni P,Jain P,Pandey S,Raghuwanshi S,bagulkar b "Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants." , ss.6 - 9, 2020. 10.14744/eej.2019.07279
ISNAD bilaiya, sakshi vd. "Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants". (2020), 6-9. https://doi.org/10.14744/eej.2019.07279
APA bilaiya s, Patni P, Jain P, Pandey S, Raghuwanshi S, bagulkar b (2020). Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. European Endodontic Journal, 5(1), 6 - 9. 10.14744/eej.2019.07279
Chicago bilaiya sakshi,Patni Pallav,Jain Pradeep,Pandey Sanket Hans,Raghuwanshi Swadhin,bagulkar bhupesh Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. European Endodontic Journal 5, no.1 (2020): 6 - 9. 10.14744/eej.2019.07279
MLA bilaiya sakshi,Patni Pallav,Jain Pradeep,Pandey Sanket Hans,Raghuwanshi Swadhin,bagulkar bhupesh Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. European Endodontic Journal, vol.5, no.1, 2020, ss.6 - 9. 10.14744/eej.2019.07279
AMA bilaiya s,Patni P,Jain P,Pandey S,Raghuwanshi S,bagulkar b Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. European Endodontic Journal. 2020; 5(1): 6 - 9. 10.14744/eej.2019.07279
Vancouver bilaiya s,Patni P,Jain P,Pandey S,Raghuwanshi S,bagulkar b Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants. European Endodontic Journal. 2020; 5(1): 6 - 9. 10.14744/eej.2019.07279
IEEE bilaiya s,Patni P,Jain P,Pandey S,Raghuwanshi S,bagulkar b "Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants." European Endodontic Journal, 5, ss.6 - 9, 2020. 10.14744/eej.2019.07279
ISNAD bilaiya, sakshi vd. "Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants". European Endodontic Journal 5/1 (2020), 6-9. https://doi.org/10.14744/eej.2019.07279