Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples

Öz:
Objectives: An automated complete blood cell count (hematology) analyzer cannot measure the white blood cell (WBC)differential of some patient sample, notably those of chemotherapy patients. This is often due to the presence of atypicalcells in the sample or matrix changes caused by drugs. A technique to use isotonic sodium chloride solution (0.9% saline) asa replacement for plasma has previously been described. The aim of this study was to evaluate to use of saline replacementas a means to resolve the matrix interference effect and achieve a reliable WBC differential count for chemotherapy patients. Methods: Samples of 29 chemotherapy patients whose WBC differential count could not be calculated using a Beck man Coulter LH-780 hematology autoanalyzer (Beckman Coulter, Inc., Brea, CA, USA) were evaluated. A peripheralblood smear was performed and the saline replacement technique was applied and the samples were then re-analyzedusing the same autoanalyzer. The WBC count and differential count of WBC of a peripheral smear and a saline-replace ment sample were compared. Results: There was no statistically significant difference between the peripheral blood smear and saline replacementautoanalyzer measurements of the WBC differential count and percent of WBC. Strong relationships in WBC (r=0.99),neutrophil (r=0.98), lymphocyte (r=0.98), monocyte (r=0.91), and eosinophil (r=0.77) counts were observed in a com parison of the peripheral blood smear and saline replacement measurements. Conclusion: A saline replacement technique may be a practical solution to resolve the difficulty of the matrix interfer ence effect seen in chemotherapy patient samples and provide a WBC differential count.
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • 1. McPherson RA, Pincus MR, editors. Henry’s Clinical Diagnosis and Management by Laboratory Methods. 22nd Ed. Philadelphia: Elsevier Saunders; 2011.
  • 2. Kim JE, Kim BR, Woo KS, Han JY. Evaluation of the leukocyte differential on a new automated flow cytometry hematology analyzer, Int J Lab Hematol 2012;34(5):547–50.
  • 3. U.S. Department of the Interior, Bureau of Reclamation, Quality Assurance Guidelines for Environmental Measurements, 1998. https://www.usbr.gov/tsc/techreferences/mands/ mands-pdfs/QualityAssuranceGuidelines_2003.pdf (Accessed May 5th 2018)
  • 4. Kealey D, Haines PJ. Analytical Chemistry, its Functions and Applications. In: Hames BD, editor. BIOS Instant Notes in Analytical Chemistry, 1st Ed. London: Taylor & Francis; 2002. p.1–4.
  • 5. Ozer H. The timing of chemotherapy-induced neutropenia and its clinical and economic impact. Oncology (Williston Park) 2006;20(5 Suppl 4):11–5.
  • 6. Dasgupta A, Sepuldeva JL. Accurate Results in the Clinical Laboratory A Guide to Error Detection and Correction. 1st Ed. St. Louis, MO: Elsevier; 2013.
  • 7. Waider KL. Rouleaux and saline replacement. Immunohematology 2018;34(3):91-2.
  • 8. Zandecki M, Genevieve F, Gerard J, Godon A. Spurious counts and spurious results on haematology analysers: a review. Part II: white blood cells, red blood cells, haemoglobin, red cell indices and reticulocytes. Int J Lab Hematol 2007;29(1):21– 41.
  • 9. COULTER LH 780 System, Reference. Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. https://www.beckmancoulter.com/download/file/wsr-32377/773021AD?type- =pdf. Accessed January 5th 2018.)
  • 10. Substantial Equivalence Determinatıon Decision Summary Assay And Instrument Combinatıon Template. https://www. accessdata.fda.gov/cdrh_docs/reviews/K081930.pdf (Accessed January 15th 2018.)
  • 11. Liu Y, Du X, Chen J, Jin Y, Peng L, Wang HHX, Luo M, Chen L, Zhao Y. Neutrophil-to-lymphocyte ratio as an independent risk factor for mortality in hospitalized patients with COVID-19. J Infect 2020;81(1):e6–12.
  • 12. Ozarda Y, Ichihara K, Bakan E, Polat H, Ozturk N, Baygutalp NK, et al. A nationwide multicentre study in Turkey for establishing reference intervals of haematological parameters with novel use of a panel of whole blood. Biochem Med (Zagreb) 2017;27(2):350–77.
  • 13. Zeng SG, Zeng TT, Jiang H, Wang LL, Tang SQ, Sun YM, et al. A simple, fast correction method of triglyceride interference in blood hemoglobin automated measurement. J Clin Lab Anal 2013;27(5):341–5.
  • 14. Er TK, Ginés MAR, ChenYT, Tsai LY, Chang JG, Lo YJ, et al. Erroneous result of white blood cell differential count in a patient with mixed hyperlipidemia, Clin Chem Lab Med 2008;46(7):1054–5.
APA BOZKURT YAVUZ H, Yaman H, TAT M, AKIDAN O, Orem A, KARAHAN S, Aliyazıcıoğlu Y (2021). Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. , 161 - 165. 10.14744/ijmb.2021.09825
Chicago BOZKURT YAVUZ Hatice,Yaman Hüseyin,TAT MUSTAFA,AKIDAN OSMAN,Orem Asım,KARAHAN Süleyman Caner,Aliyazıcıoğlu Yüksel Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. (2021): 161 - 165. 10.14744/ijmb.2021.09825
MLA BOZKURT YAVUZ Hatice,Yaman Hüseyin,TAT MUSTAFA,AKIDAN OSMAN,Orem Asım,KARAHAN Süleyman Caner,Aliyazıcıoğlu Yüksel Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. , 2021, ss.161 - 165. 10.14744/ijmb.2021.09825
AMA BOZKURT YAVUZ H,Yaman H,TAT M,AKIDAN O,Orem A,KARAHAN S,Aliyazıcıoğlu Y Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. . 2021; 161 - 165. 10.14744/ijmb.2021.09825
Vancouver BOZKURT YAVUZ H,Yaman H,TAT M,AKIDAN O,Orem A,KARAHAN S,Aliyazıcıoğlu Y Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. . 2021; 161 - 165. 10.14744/ijmb.2021.09825
IEEE BOZKURT YAVUZ H,Yaman H,TAT M,AKIDAN O,Orem A,KARAHAN S,Aliyazıcıoğlu Y "Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples." , ss.161 - 165, 2021. 10.14744/ijmb.2021.09825
ISNAD BOZKURT YAVUZ, Hatice vd. "Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples". (2021), 161-165. https://doi.org/10.14744/ijmb.2021.09825
APA BOZKURT YAVUZ H, Yaman H, TAT M, AKIDAN O, Orem A, KARAHAN S, Aliyazıcıoğlu Y (2021). Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. International Journal of Medical Biochemistry, 4(3), 161 - 165. 10.14744/ijmb.2021.09825
Chicago BOZKURT YAVUZ Hatice,Yaman Hüseyin,TAT MUSTAFA,AKIDAN OSMAN,Orem Asım,KARAHAN Süleyman Caner,Aliyazıcıoğlu Yüksel Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. International Journal of Medical Biochemistry 4, no.3 (2021): 161 - 165. 10.14744/ijmb.2021.09825
MLA BOZKURT YAVUZ Hatice,Yaman Hüseyin,TAT MUSTAFA,AKIDAN OSMAN,Orem Asım,KARAHAN Süleyman Caner,Aliyazıcıoğlu Yüksel Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. International Journal of Medical Biochemistry, vol.4, no.3, 2021, ss.161 - 165. 10.14744/ijmb.2021.09825
AMA BOZKURT YAVUZ H,Yaman H,TAT M,AKIDAN O,Orem A,KARAHAN S,Aliyazıcıoğlu Y Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. International Journal of Medical Biochemistry. 2021; 4(3): 161 - 165. 10.14744/ijmb.2021.09825
Vancouver BOZKURT YAVUZ H,Yaman H,TAT M,AKIDAN O,Orem A,KARAHAN S,Aliyazıcıoğlu Y Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. International Journal of Medical Biochemistry. 2021; 4(3): 161 - 165. 10.14744/ijmb.2021.09825
IEEE BOZKURT YAVUZ H,Yaman H,TAT M,AKIDAN O,Orem A,KARAHAN S,Aliyazıcıoğlu Y "Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples." International Journal of Medical Biochemistry, 4, ss.161 - 165, 2021. 10.14744/ijmb.2021.09825
ISNAD BOZKURT YAVUZ, Hatice vd. "Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples". International Journal of Medical Biochemistry 4/3 (2021), 161-165. https://doi.org/10.14744/ijmb.2021.09825