Yıl: 2024 Cilt: 6 Sayı: 1 Sayfa Aralığı: 4 - 10 Metin Dili: İngilizce DOI: 10.36519/idcm.2024.252.1 İndeks Tarihi: 04-04-2024

Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study

Öz:
Objective: Reuse of medical devices poses risks concerning technical issues and patient safety. In this study, we aimed to examine the structural changes in catheters that occur due to the reuse with the aid of electron microscopy. Materials and Methods: The effects of hydrogen peroxide (HP) and ethylene oxide (EO) sterilization on four percutaneous transluminal coronary angioplasty (PTCA) catheters and control PTCA catheters were examined by scanning electron microscope (SEM). Each catheter sample was divided into four parts during the SEM examination, and a total of 20 pieces were examined. Catheters were reprocessed through every regular sterilization step and used solely for the study, not in patients. Statistical evaluations of histological scoring made on images obtained from scanning electron microscopic images were made using the GraphPad Prism 8 program. Results: Electron microscopical examination showed that HP sterilization caused more robust and deeper lines compared to EO. These distortions increased directly with the increase in the reprocessing cycle. In EO, no significant damage was detected within five cycles in contrast to HP; however, the harmful effects of EO were seen over five cycles. Unprocessed samples had no damage. Outer and inner deterioration was significantly higher in the EO>5 group and HP>5 group than in the control group. However, the bacterial contamination score in the EO>5 group was higher than the control group. Conclusion: Our findings showed that HP and EO sterilizations caused some deterioration in the inner and outer surfaces of PTCA catheter samples. We recommend reprocessing using EO, the least damaging method, when necessary, and paying attention not to exceed five cycles when necessary.
Anahtar Kelime: ethylene oxide percutaneous transluminal coronary angioplasty scanning elec- tron microscopy single-use instrument

Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • 1 Rutala WA, Weber DJ. Disinfection, sterilization, and antisepsis: An overview. Am J Infect Control. 2019;47S:A3-A9. [CrossRef]
  • 2 Popp W, Rasslan O, Unahalekhaka A, et al. What is the use? An international look at reuse of single-use medical devices. Int J Hyg Environ Health. 2010;213(4):302-7. [CrossRef]
  • 3 Wang D, Wu J. Reprocessing and reuse of single-use med-ical devices in China: a pilot survey. BMC Public Health. 2019;19(1):461. [CrossRef]
  • 4 Amarante JM, Toscano CM, Pearson ML, Roth V, Jarvis WR, Levin AS. Reprocessing and reuse of single-use medical devices used during hemodynamic procedures in Brazil: a widespread and largely overlooked problem. Infect Control Hosp Epidemiol. 2008;29(9):854-8. [CrossRef]
  • 5 Perçin D. Tek kullanımlık malzemelerin cerrahide kullanımı [Internet]. In: 3. Ulusal Sterilizasyon Ameliyathane Dezenfeksiyon (SAD) Kongresi - Kongre Kitabı. [cited January 12, 2022]. Available from: https://www.das.org.tr/dosya/SAD2016-kitap- Web.pdf
  • 6 Sağlık Teknolojisi Değerlendirme Yönergesi [Internet]. Ankara: Ministry of Health. [cited June 20, 2023]. Available from: https://dosyamerkez.saglik.gov.tr/Eklenti/33489,std-yonergesipdf.pdf?0
  • 7 Zenciroğlu D. [Truth about the use of disposable equipment]. Ankem Derg. 2014;28:67-9. Turkish.
  • 8 Alfa MJ. Medical-device reprocessing. Infect Control Hosp Epidemiol. 2000;21(8):496-8. [CrossRef]
  • 9 Rutala WA, Weber DJ. Disinfection, sterilization, and antisepsis: An overview. Am J Infect Control. 2016;44(5 Suppl):e1-6. [CrossRef]
  • 10 Akcay A, Akcay M. [Determining the cost of reuse of disposable devices in cardiology]. Klimik Derg. 2019;32:136-45. Turkish. [CrossRef]
  • 11 Comis J. Reusing single-use cardiac catheters. Dimens Health Serv. 1991;68(7):8-9.
  • 12 Gerlock AJ Jr, Regen DM, Shaff MI. An examination of the physical characteristics leading to angioplasty balloon rupture. Radiology. 1982;144(2):421-2. [CrossRef]
  • 13 Frank U, Herz L, Daschner FD. Infection risk of cardiac catheterization and arterial angiography with single and multi- ple use disposable catheters. Clin Cardiol. 1988;11(11):785-7. [CrossRef]
  • 14 Jayabalan M. Sterilization and reprocessing of materials and medical devices--reusability. J Biomater Appl. 1995;10(1):97- 112. [CrossRef]
  • 15 Fagih B, Eisenberg MJ. Reuse of angioplasty catheters and risk of Creutzfeldt-Jakob disease. Am Heart J. 1999;137(6):1173-8. [CrossRef]
  • 16 Yoshida R, Kobayashi H. Influence of hydrogen peroxide sterilization on plastic surface. JHAI. 2013;6(1):19-23.
  • 17 Yoshida R, Kobayashi H. The Influence of Low-Temperature Sterilization on Plastic Surface. 14th World Sterilization Congress & 8th National Sterilization Disinfection Congress of Turkey. Ankara: Bilimsel Tıp Yayınevi; 2013.p. 220.
  • 18 Unverdorben M, Quaden R, Werner C, Bloss P, Degenhardt R, Ackermann H, et al. Change of the mechanical properties of two different balloon catheters with increasing numbers of cycles of resterilization. Catheter Cardiovasc Interv. 2003;58(1):29-33. [CrossRef]
  • 19 Gelamo RV, Sene EC, Paiva L, Oliveira Cda C, Maltos AL, Schreiner WH, et al. Effects of reprocessing on chemical and morphological properties of guide wires used in angioplasty. Rev Bras Cir Cardiovasc. 2013;28(3):331-7. [CrossRef]
  • 20 Singh J, Thingnam SK, Das D, Singh H, Sharma R, Vijayvergia R. Surgical removal of entrapped and broken percutaneous transluminal coronary angioplasty balloon catheter. Interact Cardiovasc Thorac Surg. 2007;6(1):94-6. [CrossRef]
  • 21 Butler L, Worthly LI. Reuse of flow-directed balloon-tipped catheters. Br Med J (Clin Res Ed). 1982;284(6317):707. [CrossRef]
  • 22 Chan AW, Rabinowitz A, Webb JG, Dodek A, Jarochowski M, Lohavanichbutr K, et al. Adverse events associated with balloon rupture during percutaneous coronary intervention. Can J Cardiol. 1999;15(9):962-6. English, French.
  • 23 Alfonso F, Pérez-Vizcayno MJ, Hernández R, Goicolea J, Fernández-Ortíz A, Bañuelos C, et al. Clinical and angiographic implications of balloon rupture during coronary stenting. Am J Cardiol. 1997;80(8):1077-80. [CrossRef]
  • 24 Singh B, Kaul U, Thatai D. Crimped stent-induced pin hole rupture of balloon resulting in extensive coronary dissection. J Invasive Cardiol. 1997;9(6):432-4.
  • 25 Crawford TC, Eagle KA. Reuse of catheters and devices labelled for single use: evidence, recommendations and oversight. Heart Asia. 2018;10(2):e011033. [CrossRef]
  • 26 Detenkofer M, Dascher FD. Nosocomial infections associated with cardiac catheterization and other invasive procedures in cardiology. Hospital Epidemiology and Infection Control. 3rd ed. Philadelphia: Lippincott Williams & Wilkins. 2004;117-22.
  • 27 Bourassa MG. The reuse of single-use balloon angioplasty catheters: is it now legitimate? Am J Cardiol. 1996;78(6):673-4. [CrossRef]
  • 28 Shaw JP, Eisenberg MJ, Azoulay A, Nguyen N. Reuse of catheters for percutaneous transluminal coronary angioplasty: effects on procedure time and clinical outcomes. Catheter Cardiovasc Interv. 1999;48(1):54-60. [CrossRef]
  • 29 Greene VW. Reuse of medical devices labeled for single-use. In: Mayhall GC. Hospital epidemiology and infection control. 3rd ed. Philadelphia: Lippincott Williams & Wilkins; 2004. p.1535- 45.
  • 30 Collignon PJ, Graham E, Dreimanis DE. Reuse in sterile sites of single-use medical devices: how common is this in Australia? Med J Aust. 1996;164(9):533-6. [CrossRef]
  • 31 Habib M, Musran A, Al Agha M. The reuse of (single-use) car- diac disposable of coronary angiography and angioplasty: safety and economic issues in Gaza Strip. On J Cardiovas Res. 2019;1(4). [CrossRef]
APA Kuruoglu T, Onger M, Altun G, atilla a, Esen S (2024). Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. , 4 - 10. 10.36519/idcm.2024.252.1
Chicago Kuruoglu Tuba,Onger Mehmet Emin,Altun Gamze,atilla aynur,Esen Saban Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. (2024): 4 - 10. 10.36519/idcm.2024.252.1
MLA Kuruoglu Tuba,Onger Mehmet Emin,Altun Gamze,atilla aynur,Esen Saban Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. , 2024, ss.4 - 10. 10.36519/idcm.2024.252.1
AMA Kuruoglu T,Onger M,Altun G,atilla a,Esen S Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. . 2024; 4 - 10. 10.36519/idcm.2024.252.1
Vancouver Kuruoglu T,Onger M,Altun G,atilla a,Esen S Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. . 2024; 4 - 10. 10.36519/idcm.2024.252.1
IEEE Kuruoglu T,Onger M,Altun G,atilla a,Esen S "Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study." , ss.4 - 10, 2024. 10.36519/idcm.2024.252.1
ISNAD Kuruoglu, Tuba vd. "Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study". (2024), 4-10. https://doi.org/10.36519/idcm.2024.252.1
APA Kuruoglu T, Onger M, Altun G, atilla a, Esen S (2024). Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. Infectious diseases and clinical microbiology (Online), 6(1), 4 - 10. 10.36519/idcm.2024.252.1
Chicago Kuruoglu Tuba,Onger Mehmet Emin,Altun Gamze,atilla aynur,Esen Saban Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. Infectious diseases and clinical microbiology (Online) 6, no.1 (2024): 4 - 10. 10.36519/idcm.2024.252.1
MLA Kuruoglu Tuba,Onger Mehmet Emin,Altun Gamze,atilla aynur,Esen Saban Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. Infectious diseases and clinical microbiology (Online), vol.6, no.1, 2024, ss.4 - 10. 10.36519/idcm.2024.252.1
AMA Kuruoglu T,Onger M,Altun G,atilla a,Esen S Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. Infectious diseases and clinical microbiology (Online). 2024; 6(1): 4 - 10. 10.36519/idcm.2024.252.1
Vancouver Kuruoglu T,Onger M,Altun G,atilla a,Esen S Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study. Infectious diseases and clinical microbiology (Online). 2024; 6(1): 4 - 10. 10.36519/idcm.2024.252.1
IEEE Kuruoglu T,Onger M,Altun G,atilla a,Esen S "Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study." Infectious diseases and clinical microbiology (Online), 6, ss.4 - 10, 2024. 10.36519/idcm.2024.252.1
ISNAD Kuruoglu, Tuba vd. "Detrimental Effects of Sterilization Types on Single-Use Coronary Angioplasty Catheters for Reuse: An Electron Microscopic Study". Infectious diseases and clinical microbiology (Online) 6/1 (2024), 4-10. https://doi.org/10.36519/idcm.2024.252.1